TITULO

Biomechanical Evaluation of a Novel Expandable Vertebral Augmentation System Using Human Cadaveric Vertebrae

Ming-Kai Hsieh    
Wen-Jer Chen    
Mel S. Lee    
Sheng-Yu Lin    
Mu-Yi Liu    
De-Mei Lee and Ching-Lung Tai    

Resumen

Unacceptable sagittal alignment and cement leakage are major concerns of percutaneous vertebroplasty when treating patients with painful vertebral osteoporotic compression fractures. To maintain the restored vertebral height and reduce the reliance on cement as the major stabilizer, an expandable vertebral augment system (EVA®) made of titanium alloy consisting of a rigid tube encased by a barrel with an anterior expansion mechanism was developed. The aim of the current study was to determine whether this novel design is as effective as existing procedures in terms of height restoration and biomechanical performance. Eight osteoporotic vertebrae (T12-L3) confirmed by dual-energy X-ray absorptiometry from two fresh-frozen human cadavers (70- and 72-year-old females) were used. Twenty-five percent reduced anterior wedge vertebral compression fractures were created using a material testing machine. Four randomized specimens were augmented with EVA® (Chang Gu Biotechnology Co. Ltd., Taipei city, Taiwan), and another four randomized specimens were augmented with OsseoFix® (AlphaTec Spine Inc., Carlsbad, CA, USA). The implant size and cement volume were controlled. The anterior vertebral body height (VBH) ratio and pre/postaugmented ultimate strength and stiffness were measured and compared. The mean anterior VBH restoration ratio was 8.54% in the EVA® group and 8.26% in the OsseoFix® groups. A significant difference from augmentation was measured in both groups (p < 0.05), but there was no significant difference between the EVA® and OsseoFix® groups in anterior VBH restoration. The ultimate strengths of the EVA® and OsseoFix® groups were 6071.4 ± 352.6 N and 6262.9 ± 529.2 N, respectively, both of which were statistically significantly higher than that of the intact group (4589.9 ± 474.6 N) (p < 0.05). The stiffnesses of the EVA®, OsseoFix®, and intact groups were 1087.2 ± 176.9, 1154.9 ± 168.9, and 1637.3 ± 340.8 N/mm, respectively, indicating that the stiffness was significantly higher in the intact group than in both the EVA® and OsseoFix® groups (p < 0.05). No significant differences were observed between the two augmentation procedures in height restoration or ultimate strength and stiffness. This novel EVA® system showed comparable height restoration and biomechanical performance to those of existing implants for human cadaveric osteoporotic compression fractures. Potential advantages of preventing cement posterior leakage and promoting cement interdigitation are expected with this ameliorated design.

 Artículos similares

       
 
Zanna Fiodorenko-Dumas, Ilias Dumas, Mateusz Kowal, Adrianna Machnikowska, Ewa Gieysztor, Slawomir Winiarski and Malgorzata Paprocka-Borowicz    
Background: After bilateral transfemoral amputation, people may experience limitations in everyday life due to limited mobility and prosthesis problems. Materials and method: The case study covered a 54-year-old man after bilateral traumatic amputation o... ver más
Revista: Applied Sciences

 
Guichun Wang, Jie Zhang and Xuan Kong    
Nowadays, the comfort of drivers and passengers is drawing more and more attention. The interaction between vehicles and the asphalt road can cause coupled vibration and reduce the comfort of passengers. Therefore, the research on human?vehicle?road coup... ver más
Revista: Applied Sciences

 
Bartosz Wieczorek, Mateusz Kukla, Dominik Rybarczyk and Lukasz Wargula    
Purpose: The main purpose of the research conducted was the analysis of kinematic and biomechanical parameters measured during manual wheelchair ramp-climbing with the use of the anti-rollback system and the comparison of the values tested with the manua... ver más
Revista: Applied Sciences

 
Audrey Bosquet, Caitlin Mueller and A.E. Hosoi    
Sports bras are critical to the comfort and performance of female athletes, yet mechanical models of sports bras are generally not used to guide their design. Typically, assessing any sports bra?s performance requires time-consuming and expensive biomech... ver más
Revista: Applied Sciences

 
Emilsy Medina Chacón, Ruth Illada García    
El presente artículo expone el diseño de un método heurístico para el Balance de líneas de ensamble con consideraciones ergonómicas, el cual está basado en la combinación de las teorías de balance de líneas tradicionales y dos métodos de evaluación ergon... ver más