Inicio  /  Instruments  /  Vol: 5 Par: 3 (2021)  /  Artículo
ARTÍCULO
TITULO

Pulse-Height and 2-D Charge-Spread Single-Pixels Studies on a LuYAP:Ce Scintillation Array

Raffaele Scafè    
Marco Puccini    
Rosanna Pellegrini and Roberto Pani    

Resumen

The present work deals with a 10 × 10 array of (Lu0.7Y0.3)AP:Ce 2 × 2 × 10 mm3 pixels, manufactured by Crytur (Cz), that has been utilized in a previous paper. The crystal-array has been coupled to an 8 × 8 anodes H10966 model Hamamatsu (Jp) Position-Sensitive Photo Multiplier Tube (PSPMT) connected to electronics for single events scintillation read-out. The response of such a detector has been studied under Co-57, or Ba-133, or Cs-137 gamma-ray emissions, as well as with Lu-176 self-activity only. The present work is aimed at characterizing the individual crystal-pixels? single-event responses in terms of pulse-height and of spreads of the 2-D charge-distributions. In particular, the charge-spread characterization pointed out several defects in the crystal-array assembly, not detected by usual pulse-height studies. The diagnostic method based on charge-spread analysis seems also well suited for scintillation array characterizations for gamma-ray detectors studies, as well as for quality controls of such pixelated devices during the lifetime of systems in the field of radionuclide medical imaging (SPECT and PET). The method is also appropriate for other applications where gamma-ray spectroscopy is required, like nuclear physics, astrophysics, astroparticle physics, homeland security, and non-proliferation.