Enhanced Electrical Injury Using Triangular Interdigitated Electrodes for Catheter-Based Irreversible Electroporation

1
Inha Research Institute for Aerospace Medicine, Inha University, Incheon 22212, Republic of Korea
2
Center for Sensor Systems, Inha University, Incheon 22212, Republic of Korea
3
Department of Electrical and Computer Engineering, College of Engineering, Inha University, Incheon 22212, Republic of Korea
*
Author to whom correspondence should be addressed.
Appl. Sci. 2023, 13(14), 8455; https://doi.org/10.3390/app13148455
Submission received: 17 May 2023 / Revised: 12 June 2023 / Accepted: 13 June 2023 / Published: 21 July 2023
(This article belongs to the Section Electrical, Electronics and Communications Engineering)

Abstract

:

Featured Application

The triangular interdigitated electrodes proposed in this paper could potentially provide an alternative strategy for designing interdigitated electrodes for catheter-based irreversible electroporation.

Abstract

Irreversible electroporation (IRE) is a promising nonthermal ablation technique that uses high-voltage electrical pulses to create permanent pores in the cell membrane of target tissue. Recently, endoscopic IRE with catheter-based electrodes has attracted significant attention as a potential alternative tool for gastrointestinal tumors, but it has been challenged owing to the limited electric field distribution in an in-plane electrode configuration, in which rectangular interdigitated electrodes (IDEs) are commonly used. Herein, we report an enhanced electrical injury in tissue using triangular IDEs that cause strong electric fields to be induced at the tip of the electrode fingers. A set of 10 pulses with a duration of 100 μs and a frequency of 1 Hz were delivered to the tissue, and a finite element method was used to calculate the electrical injury in the gastrointestinal model. The probability of cell death by electrical injury at the triangular IDEs increases by approximately 10 times compared to that of conventional rectangular IDEs at the same electrode distance. These results could potentially pave the way toward designing electrodes in catheter-based IRE devices.

1. Introduction

Irreversible electroporation (IRE) is a nonthermal ablation technique that uses high-voltage electrical pulses to create permanent pores in the cell membrane of target tissue, leading to cell death [1,2]. Currently, IRE is considered one of the most efficient methods of destroying tumors in various organs, including the liver, pancreas, kidney, lung, brain, and prostate [3,4,5,6,7,8,9,10,11,12]. In clinical use, needle-shaped electrodes puncture the target tissue surrounding the tumor to deliver electrical pulses with the electric field strength of a few kV/cm [13]. However, this electrode configuration is unsuitable for vulnerable gastrointestinal tracts with several hollow organs [14].
Catheter-based IRE presents a promising alternative tool for effective treatment of gastrointestinal tumors or undesired tissues [14,15]. In the clinical process, a catheter is inserted into the body and guided to the tumor site, where it delivers electrical pulses to the tumor. This method is suitable for preventing perforation or injury. Jeon et al. designed two types of catheters, the needle-type and the basket-type, and investigated the feasibility and effectiveness of endoscopic IRE therapy on hollow viscous tissues [16]. Further, Rohan et al. demonstrated the feasibility of IRE based on a three-electrode tubular catheter in an experimental liver model in both ex vivo and in vivo studies [17]. In addition, interdigitated electrodes (IDEs) have been studied for their ability to reduce the required voltage and localize the electric fields at the desired region [13,14]. IDEs further allow the electrode distance to be closer and thinner due to simple in-plane geometry [18,19]. Most of the finger structures of IDEs reported so far have a rectangular cross-section, and most studies in this area have focused on the effects of the finger’s width, length, and period have been conducted [20,21]. Similarly, studies on IDEs used for electroporation have mainly concentrated on electrodes with a rectangular cross-section and various pulse settings (duration, total pulse number, frequency, voltage), but there has been no study on finger shape according to the authors’ knowledge [13,22,23,24]. Therefore, further improvement in IDEs to enhance the electric field strength is required to increase the practicability of catheter-based IRE.
In this work, we propose triangular IDEs for catheter-based IRE by demonstrating electrical injury in a gastrointestinal model. The triangular IDEs allow strong electric fields to be induced at the tip of the electrode fingers and are simply fabricated using laser-induced graphene technology on a flexible polyimide film [18]. A three-dimensional finite element model based on COMSOL Multiphysics 5.4 software was adopted to investigate the enhancement in the electric field through the electrode design. With triangular IDEs, the probability of cell death due to electrical injury (EI) in tissue increased around 10 times compared to that of conventional rectangular IDEs at the same electrode distance. These results could indicate an alternative strategy for designing IDEs for catheter-based IRE.

2. Materials and Methods

2.1. Model of Gastrointestinal Tract

The gastrointestinal tract model, which included the mucosal layer (ML), submucosal layer (SL), and muscularis propria (MP), had a length of 40 mm and a thickness of 4 mm, as shown in Figure 1. The thicknesses of the ML, SL, and MP were 1, 1, and 2 mm, respectively. Each layer was considered to have isotropic and homogeneous electrical and thermal properties as reported in a previous study [14,25,26,27].

2.2. Electrode Configuration

The basic concept of the electrode design is shown in Figure 1. To compare the enhancement in the electric field, we designed two IDE configurations: (1) a conventional rectangular pattern and (2) a triangular pattern. The anode and cathode, which were both located inside the gastrointestinal track, both had a thickness of 0.2 mm.

2.3. Numerical Analysis

A three-dimensional finite element model of IRE was developed using COMSOL Multiphysics 5.4 software (COMSOL Inc., Palo Alto, CA, USA), which was used to investigate the electric field distribution created by the voltage applied to the electrodes and the resulting electromagnetic heating from current flow. The electric field was determined by solving the governing Laplace equation [14]:
σ φ = 0 ,
where φ is the voltage (V) and σ is the electrical conductivity of the tissue (S/m).
The temperature distribution was solved using the modified Pennes’ bioheat transfer equation [14,28]:
k T + ω b c b T a T + Q m + σ E 2 = ρ c t T t ,  
where k is the thermal conductivity of the tissue ( W / m K ) , T is the temperature ( K ), ω b is the blood perfusion rate ( 1 / s ), c b is the specific heat capacity of blood ( J / k g K ), T a is the arterial temperature ( K ), Q m is the metabolic heat generation in the tissue ( W / m 3 ), ρ is the tissue density ( k g / m 3 ), and c t is the specific heat capacity of the tissue ( J / k g K ). The joule heating term ( σ E 2 ) was averaged over the entire duration of the treatment instead of calculating the heating from each individual pulse.
The tissue electrical conductivity was determined using the following equation to incorporate changes occurring due to electroporation and the thermal effect [14]:
σ d E , T = σ 0 1 + A f l c 2 h s n o r m E E d e l t a , E r a n g e + α T T 0 ,
where σ 0 is the baseline electrical conductivity, A is the multiplication factor, α is the conductivity–temperature coefficient, T 0 is the physiological temperature, and E d e l t a is the threshold electric field in electroporation. f l c 2 h s is a smoothed Heaviside function defined in COMSOL software, which changes from zero to one over the range of ± E r a n g e . A , E r a n g e , and E d e l t a are given in Table 1. The conductivity–temperature coefficient was taken as 0.02 ( 1 / K ) in this study. The value of 0.02 (1/K) was chosen from the literature [29,30].

2.4. Probability of Cell Death due to EI

The probability of cell death by EI during IRE was calculated using the Peleg–Fermi model, as follows [14]:
S = 1 1 + e E E c n   A ( n )
where S is the proportion of surviving cells after IRE and E is the electric field strength. E c ( n ) and A ( n ) are two functions that depend on the number of pulses:
E c n = E 0 exp   ( k 1 n )
A n = A 0 exp   ( k 2 n )
where E 0 , A 0 , k 1 , and k 2 are the regression coefficients that are related to the pulse duration and cell type, having values of 3996 V / c m , 1441 V / c m , 0.03, and 0.06, respectively, as described in the literature [35]. The probability of cell death due to EI can be calculated as follows:
P r o b a b i l i t y   o f   E I   ( P E I ) = 1 S .
Thus, a certain region where the PEI is greater than 0.90 is determined to be treated with IRE.

3. Results

The simulation models of the gastrointestinal tract and triangular IDEs are illustrated in Figure 1. The gastrointestinal tract was composed of the mucosal layer (ML), submucosal layer (SL), and muscularis propria (MP). The triangular IDEs were located inside the gastrointestinal tract. The electrical and thermal properties of gastrointestinal tract were based on the literature, and the electrode was set to be gold in the COMSOL built-in library (Table 1).
The electrical conductivity response according to the change in the temperature and electric field of the tissue is shown in Figure 2. The electrical conductivity was calculated through Eq. 3 within the temperature range of 37–57 °C and the electric field range of 100–800 V/cm. The ranges of electrical conductivity change for the ML, SL, and MP used in this study were 0.511–1.638, 0.251–0.803, and 0.202–0.586 S/m, respectively. The above values were used in FEM simulation for the electric field distribution in the tissue and for probability of cell death due to EI (PEI). In addition, it should be noted that the influence of temperature on conductivity was small enough to be negligible compared with the change in electrical conductivity for electric field strength [33].
In order to investigate the effect of triangular IDEs, the electrical performances of rectangular and triangular IDEs with an identical electrode distance were compared. The rectangular IDEs had a finger length of 15 mm, finger width of 5 mm, electrode distance of 5 mm, and three fingers, as shown in Figure 3a. Two triangular IDEs were designed with (1) three fingers with a finger length of 15 mm, a base length of 5 mm, and a ~2 times longer electrode distance (d ≈ 11 mm) than rectangular IDEs as shown in Figure 3b; and (2) five fingers with a finger length of 15 mm, a base length of 5 mm, and an identical electrode distance (d ≈ 5 mm) of rectangular IDES as shown in Figure 3c. The voltage distribution for each electrode configuration is shown in Figure 3. A set of 10 pulses with a duration of 100 μs and a frequency of 1 Hz were delivered to the tissue, and the voltage of each pulse was set to be 1 kV. As shown in the voltage profile of the rectangular (d = 5 mm) and the triangular IDEs (d ≈ 11 mm, d ≈ 5 mm) at x = 20 mm and z = 3.5 mm in Figure 3d, the slopes of the rectangular IDEs (d = 5 mm) and the triangular IDEs (d ≈ 5 mm) in the region shaded in green were similar at 159.8 and 145.7, respectively, but the slope of the triangular IDEs (d ≈ 11 mm) was 83.2, showing a large difference.
The electric field distribution in the rectangular and triangular IDEs was evaluated as shown in Figure 4. The top panel shows the electric field distribution in the x-y plane view and the bottom panel shows the y–z cross-sectional view. The electric field was strongly focused on the edge of the rectangular IDE (d = 5 mm), and the maximum electric field was about 1.09 × 106 V/m. For the triangular IDE, a strong electric field was observed at the tip region, and the maximum magnitudes were about 2.54 ×106 V/m at d ≈ 11 mm and 3.17 × 106 V/m at d ≈ 5 mm. Compared to the triangular IDEs with a d ≈ 11 mm, the triangular IDEs with a d ≈ 5 mm had a smaller effective electrode spacing, so the magnitude of the electric field appears larger.
Figure 5 shows the distribution of PEI calculated from Equation (7) for the rectangular IDE (d = 5 mm), triangular IDE with three fingers (d ≈ 11 mm), and triangular IDE with five fingers (d ≈ 5 mm). From these results, a relative area ratio of PEI > 0.90 was calculated. The relative area ratio of PEI was defined as the ratio of the valid area of PEI > 0.90 to the total cross-sectional area. Figure 6 shows a comparison of the relative area ratio of PEI in rectangular IDEs (d = 5 mm), triangular IDEs with three fingers (d ≈ 11 mm), and triangular IDEs with five fingers (d ≈ 5 mm). The relative area ratios of the rectangular IDE (d = 5 mm), triangular IDE with three fingers (d ≈ 11 mm), and triangular IDE with five fingers (d ≈ 5 mm) were 0.18%, 1.45%, and 1.97%, respectively. The relative area ratio of the triangular IDEs with five fingers increased around 10 times compared to that of the conventional rectangular IDEs at the same electrode distance.

4. Discussion

Catheter-based IRE is a promising tool for effective treatment of gastrointestinal tumors, and various electrode configurations, including needle-type, basket-type, and interdigitated electrodes, have been studied. Among them, IDEs have the advantages of allowing for thinner and smaller electrodes and forming a high electric field at a low voltage. Although several studies have been performed to evaluate IRE characteristics according to the design parameters of rectangular IDEs, relatively few studies have been conducted on the electrode configuration of IDEs.
Triangular IDEs allow strong electric fields to be induced at the tip of the electrode fingers, and can deliver electric field energy to the tissue more efficiently than conventional rectangular IDEs. Therefore, triangular IDEs can provide an alternative catheter-based IRE electrode design.
In general, about 70–100 pulses are needed as a threshold for almost-complete tissue ablation for IRE [14,36]. However, due to the limitation of computation capacity, we only delivered a set of 10 pulses with a duration of 100 μs and a frequency of 1 Hz to the tissue. These simulation conditions act as a limitation in examining overall tissue ablation by IRE; however, it is deemed sufficient to compare the electric field distribution according to the finger shape of the IDEs and provides a new alternative in the electrode design of catheter-based IRE.

5. Conclusions

We propose and demonstrate triangular IDEs that can deliver electric field energy more efficiently than conventional rectangular IDEs and cause injury to a desired tissue. A strong electric field was induced at the tip of the triangular IDEs. The relative area ratio of the triangular IDEs increased by approximately 10 times compared to that of conventional rectangular IDEs at the same electrode distance. We believe that these results will be very helpful for designing electrodes in catheter-based IRE devices.

Author Contributions

Conceptualization, D.-J.L. and D.Y.K.; methodology, D.-J.L.; writing—original draft preparation, D.-J.L.; writing—review and editing, D.-J.L. and D.Y.K. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the INHA UNIVERSITY Research Grant.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Aycock, K.N.; Davalos, R.V. Irreversible Electroporation: Background, Theory, and Review of Recent Developments in Clinical Oncology. Bioelectricity 2019, 1, 214–234. [Google Scholar] [CrossRef]
  2. Wagstaff, P.G.K.; Buijs, M.; van den Bos, W.; de Bruin, D.M.; Zondervan, P.J.; de la Rosette, J.; Pes, M.P.L. Irreversible electroporation: State of the art. OncoTargets Ther. 2016, 9, 2437–2446. [Google Scholar] [CrossRef] [Green Version]
  3. Zhu, T.Y.; Ai, J.; Nie, C.H.; Zhou, G.H.; Chen, X.H.; Zhang, Y.L.; Zhou, T.Y.; Chen, S.Q.; Wang, B.Q.; Zheng, S.S.; et al. Feasibility of computed tomography-guided irreversible electroporation for porcine kidney ablation. J. Cancer Res. Ther. 2020, 16, 1125–1128. [Google Scholar] [CrossRef] [PubMed]
  4. Yu, M.L.; Li, S. Irreversible electroporation for liver cancer ablation: A meta analysis. Eur. J. Surg. Oncol. 2022, 48, 1321–1330. [Google Scholar] [CrossRef] [PubMed]
  5. Wagstaff, P.G.K.; de Bruin, D.M.; van den Bos, W.; Ingels, A.; van Gemert, M.J.C.; Zondervan, P.J.; Verdaasdonk, R.M.; van Lienden, K.P.; van Leeuwen, T.G.; de la Rosette, J.; et al. Irreversible electroporation of the porcine kidney: Temperature development and distribution. Urol. Oncol.-Semin. Orig. Investig. 2015, 33, 7. [Google Scholar] [CrossRef]
  6. Saini, A.; Breen, I.; Alzubaidi, S.; Pershad, Y.; Sheth, R.; Naidu, S.; Knuttinen, M.G.; Albadawi, H.; Oklu, R. Irreversible Electroporation in Liver Cancers and Whole Organ Engineering. J. Clin. Med. 2019, 8, 11. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Ong, S.; Leonardo, M.; Chengodu, T.; Bagguley, D.; Lawrentschuk, N. Irreversible Electroporation for Prostate Cancer. Life 2021, 11, 490. [Google Scholar] [CrossRef]
  8. Murphy, K.R.; Aycock, K.N.; Hay, A.N.; Rossmeisl, J.H.; Davalos, R.V.; Dervisis, N.G. High-frequency irreversible electroporation brain tumor ablation: Exploring the dynamics of cell death and recovery. Bioelectrochemistry 2022, 144, 8. [Google Scholar] [CrossRef]
  9. Morozov, A.; Taratkin, M.; Barret, E.; Singla, N.; Bezrukov, E.; Chinenov, D.; Enikeev, M.; Rivas, J.G.; Shpikina, A.; Enikeev, D. A systematic review of irreversible electroporation in localised prostate cancer treatment. Andrologia 2020, 52, 13. [Google Scholar] [CrossRef]
  10. Lv, Y.P.; Tang, X.; Peng, W.C.; Cheng, X.; Chen, S.; Yao, C.G. Analysis on reversible/irreversible electroporation region in lung adenocarcinoma cell model in vitro with electric pulses delivered by needle electrodes. Phys. Med. Biol. 2020, 65, 9. [Google Scholar] [CrossRef]
  11. Lee, E.W.; Shahrouki, P.; Peterson, S.; Tafti, B.A.; Ding, P.X.; Kee, S.T. Safety of Irreversible Electroporation Ablation of the Pancreas. Pancreas 2021, 50, 1281–1286. [Google Scholar] [CrossRef]
  12. Fang, Z.; Chen, L.C.; Moser, M.A.J.; Zhang, W.J.; Qin, Z.Y.; Zhang, B. Electroporation-Based Therapy for Brain Tumors: A Review. J. Biomech. Eng.-Trans. Asme 2021, 143, 14. [Google Scholar] [CrossRef] [PubMed]
  13. Kurata, K.; Yoshimatsu, S.; Takamatsu, H. Low-Voltage Irreversible Electroporation Using a Comb-Shaped Contact Electrode. IEEE Trans. Biomed. Eng. 2020, 67, 420–427. [Google Scholar] [CrossRef]
  14. Ren, F.; Li, Q.; Gao, X.; Zhu, K.; Zhang, J.; Chen, X.; Yan, X.; Chu, D.; Hu, L.; Gao, Z.; et al. Electrical and thermal analyses of catheter-based irreversible electroporation of digestive tract. Int. J. Hyperth. 2019, 36, 854–867. [Google Scholar] [CrossRef] [Green Version]
  15. Lee, J.M.; Choi, H.S.; Chun, H.J.; Kim, E.S.; Keum, B.; Seo, Y.S.; Jeen, Y.T.; Lee, H.S.; Um, S.H.; Kim, C.D.; et al. EUS-guided irreversible electroporation using endoscopic needle-electrode in porcine pancreas. Surg. Endosc. 2019, 33, 658–662. [Google Scholar] [CrossRef] [PubMed]
  16. Jeon, H.J.; Choi, H.S.; Keum, B.; Bang, E.J.; Lee, K.W.; Kim, S.H.; Yim, S.Y.; Lee, J.M.; Kim, E.S.; Seo, Y.S.; et al. Feasibility and effectiveness of endoscopic irreversible electroporation for the upper gastrointestinal tract: An experimental animal study. Sci. Rep. 2021, 11, 15353. [Google Scholar] [CrossRef] [PubMed]
  17. Rohan, T.; Andrasina, T.; Juza, T.; Matkulcik, P.; Cervinka, D.; Svobodova, I.; Novotna, V.; Bernard, V.; Valek, V.; Goldberg, S.N. Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter. Int. J. Hyperth. 2021, 38, 393–401. [Google Scholar] [CrossRef] [PubMed]
  18. Kim, J.; Wi, S.M.; Ahn, J.G.; Son, S.; Lim, H.; Park, Y.; Eun, H.J.; Park, J.B.; Lim, H.; Pak, S.; et al. Engineering Geometric Electrodes for Electric Field-Enhanced High-Performance Flexible In-Plane Micro-Supercapacitors. Energy Environ. Mater. 2023, e12581. [Google Scholar] [CrossRef]
  19. Choi, S.E.; Khoo, H.; Hur, S.C. Recent Advances in Microscale Electroporation. Chem. Rev. 2022, 122, 11247–11286. [Google Scholar] [CrossRef]
  20. Kosri, E.; Ibrahim, F.; Thiha, A.; Madou, M. Micro and Nano Interdigitated Electrode Array (IDEA)-Based MEMS/NEMS as Electrochemical Transducers: A Review. Nanomaterials 2022, 12, 4171. [Google Scholar] [CrossRef]
  21. Barnes, E.O.; Fernández-la-Villa, A.; Pozo-Ayuso, D.F.; Castaño-Alvarez, M.; Lewis, G.E.M.; Dale, S.E.C.; Marken, F.; Compton, R.G. Interdigitated ring electrodes: Theory and experiment. J. Electroanal. Chem. 2013, 709, 57–64. [Google Scholar] [CrossRef] [Green Version]
  22. Xu, M.; Hong, W.; Song, Z.; Guo, Z.; Wang, L.; You, M.; Lin, Z.; Yang, B.; Qin, M.; Liu, J. Multifunctional Microelectrode for Low-Voltage Pulsed Field Ablation in Atrial Fibrillation. IEEE Electron Device Lett. 2022, 43, 1965–1968. [Google Scholar] [CrossRef]
  23. Yoshimatsu, S.; Yoshida, M.; Kurata, K.; Takamatsu, H. Development of contact irreversible electroporation using a comb-shaped miniature electrode. J. Therm. Sci. Technol. 2017, 12, JTST0023. [Google Scholar] [CrossRef] [Green Version]
  24. Novickij, V.; Tabasnikov, A.; Smith, S.; Grainys, A.; Novickij, J. Analysis of Planar Circular Interdigitated Electrodes for Electroporation. Iete Tech. Rev. 2015, 32, 196–203. [Google Scholar] [CrossRef] [Green Version]
  25. Abu-Ghanem, Y.; Meydan, C.; Segev, L.; Rubin, M.; Blumenfeld, O.; Spivak, H. Gastric Wall Thickness and the Choice of Linear Staples in Laparoscopic Sleeve Gastrectomy: Challenging Conventional Concepts. Obes. Surg. 2017, 27, 837–843. [Google Scholar] [CrossRef]
  26. Yazar, F.M.; Baykara, M.; Karaagac, M.; Bulbuloglu, E. The Role of Conventional Ultrasonography in the Evaluation of Antrum Wall Thickness in Obese Patients. Obes. Surg. 2016, 26, 2995–3000. [Google Scholar] [CrossRef] [PubMed]
  27. Rawlins, L.; Rawlins, M.P.; Teel, D. Human tissue thickness measurements from excised sleeve gastrectomy specimens. Surg. Endosc. Other Interv. Tech. 2014, 28, 811–814. [Google Scholar] [CrossRef] [PubMed]
  28. Nickfarjam, A.; Firoozabadi, S.M.P. Parametric study of irreversible electroporation with different needle electrodes: Electrical and thermal analysis. Int. J. Hyperth. 2014, 30, 335–347. [Google Scholar] [CrossRef] [PubMed]
  29. Duck, F.A. Chapter 6—Electrical Properties of Tissue. In Physical Properties of Tissues; Duck, F.A., Ed.; Academic Press: Cambridge, MA, USA, 1990; pp. 167–223. [Google Scholar]
  30. Zhao, Y.; Zheng, S.; Beitel-White, N.; Liu, H.; Yao, C.; Davalos, R.V. Development of a Multi-Pulse Conductivity Model for Liver Tissue Treated With Pulsed Electric Fields. Front. Bioeng. Biotechnol. 2020, 8, 396. [Google Scholar] [CrossRef]
  31. Beitel-White, N.; Bhonsle, S.; Martin, R.C.G.; Davalos, R.V. Electrical Characterization of Human Biological Tissue for Irreversible Electroporation Treatments. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2018, 2018, 4170–4173. [Google Scholar] [CrossRef]
  32. Golberg, A.; Bruinsma, B.G.; Uygun, B.E.; Yarmush, M.L. Tissue heterogeneity in structure and conductivity contribute to cell survival during irreversible electroporation ablation by “electric field sinks’”. Sci. Rep. 2015, 5, 8485. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  33. Garcia, P.A.; Davalos, R.V.; Miklavcic, D. A Numerical Investigation of the Electric and Thermal Cell Kill Distributions in Electroporation-Based Therapies in Tissue. PLoS ONE 2014, 9, e103083. [Google Scholar] [CrossRef] [PubMed]
  34. Yang, Y.; Moser, M.; Zhang, E.; Zhang, W.; Zhang, B. Optimization of Electrode Configuration and Pulse Strength in Irreversible Electroporation for Large Ablation Volumes Without Thermal Damage. J. Eng. Sci. Med. Diagn. Ther. 2018, 1, 21002. [Google Scholar] [CrossRef]
  35. Canatella, P.J.; Karr, J.F.; Petros, J.A.; Prausnitz, M.R. Quantitative study of electroporation-mediated molecular uptake and cell viability. Biophys. J. 2001, 80, 755–764. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  36. Lindelauf, K.H.K.; Baragona, M.; Baumann, M.; Maessen, R.T.H.; Ritter, A. Pulse Parameters and Thresholds for (ir)Reversible Electroporation on Hepatocellular Carcinoma Cells in Vitro. Technol. Cancer Res. Treat. 2023, 22, 15330338221136694. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Model of gastrointestinal tract and triangular interdigitated electrodes (IDEs). The gastrointestinal track contains the mucosal layer, submucosal layer, and muscularis propria. Triangular IDEs are located on the outermost part of the balloon catheter and are in contact with the gastrointestinal track.
Figure 1. Model of gastrointestinal tract and triangular interdigitated electrodes (IDEs). The gastrointestinal track contains the mucosal layer, submucosal layer, and muscularis propria. Triangular IDEs are located on the outermost part of the balloon catheter and are in contact with the gastrointestinal track.
Applsci 13 08455 g001
Figure 2. Electrical conductivity response of the (a) mucosal layer, (b) submucosal layer, and (c) muscularis propria during irreversible electroporation (IRE).
Figure 2. Electrical conductivity response of the (a) mucosal layer, (b) submucosal layer, and (c) muscularis propria during irreversible electroporation (IRE).
Applsci 13 08455 g002
Figure 3. Voltage distribution of (a) rectangular IDEs with three fingers and electrode distance of 5 mm (d = 5 mm), (b) triangular IDEs with three fingers and electrode distance of 11 mm (d ≈ 11 mm), and (c) triangular IDEs with five fingers and electrode distance of 5 mm (d ≈ 5 mm) during IRE. The white arrow denotes the distance between electrodes. (d) The voltage variation in the rectangular (d = 5 mm) and the triangular IDEs (d ≈ 11 mm, d ≈ 5 mm) on y-axis at x = 20 mm and z = 3.5 mm.
Figure 3. Voltage distribution of (a) rectangular IDEs with three fingers and electrode distance of 5 mm (d = 5 mm), (b) triangular IDEs with three fingers and electrode distance of 11 mm (d ≈ 11 mm), and (c) triangular IDEs with five fingers and electrode distance of 5 mm (d ≈ 5 mm) during IRE. The white arrow denotes the distance between electrodes. (d) The voltage variation in the rectangular (d = 5 mm) and the triangular IDEs (d ≈ 11 mm, d ≈ 5 mm) on y-axis at x = 20 mm and z = 3.5 mm.
Applsci 13 08455 g003
Figure 4. Effects of electrode configuration and distance on the distribution of electric fields in the x–y plane view (top) and y–z cross-sectional view (bottom). The white dashed line in the top view indicates the cutting region for the cross-sectional view.
Figure 4. Effects of electrode configuration and distance on the distribution of electric fields in the x–y plane view (top) and y–z cross-sectional view (bottom). The white dashed line in the top view indicates the cutting region for the cross-sectional view.
Applsci 13 08455 g004
Figure 5. Probability of electrical injury of (a) rectangular IDEs (d = 5 mm), (b) triangular IDEs with three fingers (d ≈ 11 mm), and (c) triangular IDEs with five fingers (d ≈ 5 mm) during IRE.
Figure 5. Probability of electrical injury of (a) rectangular IDEs (d = 5 mm), (b) triangular IDEs with three fingers (d ≈ 11 mm), and (c) triangular IDEs with five fingers (d ≈ 5 mm) during IRE.
Applsci 13 08455 g005
Figure 6. Comparison of the relative area ratio of PEI > 0.90 in rectangular IDEs (d = 5 mm), triangular IDEs with three fingers (d ≈ 11 mm), and triangular IDEs with five fingers (d ≈ 5 mm).
Figure 6. Comparison of the relative area ratio of PEI > 0.90 in rectangular IDEs (d = 5 mm), triangular IDEs with three fingers (d ≈ 11 mm), and triangular IDEs with five fingers (d ≈ 5 mm).
Applsci 13 08455 g006
Table 1. Electrical and thermal properties of the gastrointestinal tract and electrode.
Table 1. Electrical and thermal properties of the gastrointestinal tract and electrode.
MaterialMuscularis PropriaSubmucosal Layer Mucosal LayerElectrode
Electrical conductivity (S/m) 0.2020.2510.5114.5 × 107
A 1.51.81.8-
E d e l t a   ( V / c m ) 500600600-
E r a n g e   ( V / c m ) 300150150-
Density ( k g / m 3 )10901027108819,300
Specific heat capacity ( J / k g K )342123723690129
Thermal conductivity ( W / m K ) 0.490.390.53317
References[14,31,32,33,34][14,31,32,33,34][14,31,32,33,34]-
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Lee, D.-J.; Kim, D.Y. Enhanced Electrical Injury Using Triangular Interdigitated Electrodes for Catheter-Based Irreversible Electroporation. Appl. Sci. 2023, 13, 8455. https://doi.org/10.3390/app13148455

AMA Style

Lee D-J, Kim DY. Enhanced Electrical Injury Using Triangular Interdigitated Electrodes for Catheter-Based Irreversible Electroporation. Applied Sciences. 2023; 13(14):8455. https://doi.org/10.3390/app13148455

Chicago/Turabian Style

Lee, Dong-Jin, and Dae Yu Kim. 2023. "Enhanced Electrical Injury Using Triangular Interdigitated Electrodes for Catheter-Based Irreversible Electroporation" Applied Sciences 13, no. 14: 8455. https://doi.org/10.3390/app13148455

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop